hemodialysis

Completion requirements

- Care for the dialysis machine

  

Steps  :

Wearing personal protective equipment.

Remove the solution stick from the jug and place it in the machine.

Connecting the dialysis tubes to the machine

·        Ensure that all doors (pump, blood reservoir) are closed to create a closed loop Rinse

·        Enter the program first.

·        Entering the machine into the Dis infection program.

· The necessary solution for the machine is a chlorine solution for acetate washing and a citric solution for bicarbonate washing.

·        Shut down the machine after it completes the program

Wipe the outside of the machine with diluted bleach at a ratio of 100 cm of bleach per liter of water for the machine body.

Wipe the parts of all control switches with 95% alcohol.

 

Protocol for sterilizing the Fresenius dialysis machine model 4008

 

First: In the case of washing using a solution:أسيتات

A rinse is done between each session.At the end of the day (every day), one of the two methods is followed.

First way :

Chemical cold disinfection PGM5 using a 10% diluted sodium hypochlorite.
The commercial name (Javel or Clorox)
The hot rinse program is working.برنامج الشطف الساخن

Second way :

Conduct a hot chemical decontamination on PGM2 (from the back of the yellow link) using one of the following media –


Fresenius Diasteril


Fresenius Citrosteril