- Nasal Continuous Positive Airway Pressure (Nasal CPAP)
Nasal CPAP is a non-invasive, safe ventilatory support
system that does not require an endotracheal tube. It delivers a continuous
positive pressure of oxygen-enriched air to keep alveoli open and prevent
collapse during exhalation. It can be delivered via mask, nasopharyngeal
catheter, or endotracheal tube, though nasal CPAP is the most commonly used.
Indications
- Respiratory distress from Hyaline
Membrane Disease in preterm neonates due to surfactant deficiency, to
improve lung efficiency and reduce the need for invasive ventilation.
- PaO₂ < 50 mmHg despite
oxygen therapy ≥60% using simpler methods.
- Moderate respiratory distress:
tachypnea (>60 breaths/min), grunting, or subcostal retractions, such
as:
- Transient neonatal tachypnea
- Meconium aspiration
- Apnea of prematurity in low
birth weight infants.
- Post-weaning from mechanical
ventilation.
Contraindications
- Respiratory failure requiring
invasive ventilation.
- Persistent apnea with
bradycardia unresponsive to CPAP.
- Upper airway malformations
(cleft palate, nasal obstruction).
- Diaphragmatic hernia.
- Severe, unstable cardiac
disease.
Nasal CPAP Components
- Gas Source: Provides a humidified mixture of oxygen and compressed
air.
- Pressure Generator: Creates continuous positive airway pressure.
- Interface: Connects the circuit to the neonate’s airway (nasal
prongs or mask).
Circuit Components
- Gas Delivery Circuit:
- Oxygen and compressed air
inlets.
- Oxygen blender to mix oxygen
and air to the required FiO₂.
- Flowmeter (8–10 L/min) and
manometer to monitor pressure.
- Humidifier: Heats and humidifies the inspired gas with a
temperature sensor.
- Pressure Generator: Maintains continuous positive pressure at the end of
exhalation to keep alveoli open.

Positive Pressure Generator (CPAP
Generator)
Other Methods to
Generate Positive Pressure (Bubble–CPAP)
- Immerse the free end of the
exhalation tube under the water surface to a
depth of 5 cm to generate the desired positive pressure in the
alveoli.
- Attach a graduated measuring
tube to the outer wall of the
sterile water bottle.
- Fill the bottle with sterile
water until the water level reaches 7
on the measuring tube.
- Immerse the tube end under the water to 5 cm, leaving a 2 cm gap
between the tube end and the bottom of the bottle to prevent excessive
alveolar pressure.
- Ensure the appearance of air
bubbles after connecting the neonate
and starting the CPAP device.
- Connect the device’s exhalation
tube to a valve.

Nasal Prongs (Double-Prong
Catheter)
- Connection: Attaches to the end of the generator to deliver the
inhaled gas circuit to the infant.
Device Connections
- Short Corrugated Tube: Connects the flow meter (on the oxygen blender) to the
humidifier.
- Inhalation Tube: A long corrugated tube carrying the humidified inhaled
gas from the humidifier to the positive pressure generator and the
double-prong nasal tube, delivering warm, moist air to the infant.
- Exhalation Tube: A corrugated tube connecting the double-prong nasal
tube to the positive pressure generator at the back, connected to a
manometer to monitor the generated pressure.

(Bubble CPAP)
Nurse’s Role:
The device must always be ready for use at any time in the
neonatal care units, with all necessary connections available, including all
sizes of double-prong nasal catheters according to the infant’s weight and
gestational age.
Preparing the Infant for Connection
to the Device
- Infection Control: Follow strict infection control measures to prevent
respiratory infections and ensure the device functions efficiently:
- Wash hands thoroughly before
handling the device, catheter, or infant.
- Wear sterile gloves during
airway suctioning.
- Keep all tubing off the floor
or away from contamination sources.
- Device Setup:
- Connect the oxygen and air
supply to the oxygen blender, adjusting the oxygen flow rate to 8–10
L/min.
- Set the oxygen concentration
to the required percentage.
- Connect the short corrugated
tube to the oxygen humidifier and insert the temperature probe into the
designated port.
- Fill the humidifier with
sterile water to the marked level, turn on the device, and set the
temperature to 37°C.
- Select the appropriate cap
size for the infant’s head.
- Select the correct nasal
catheter size by measuring the infant’s nostrils; proper fit is crucial:
- Too small → leakage,
difficult to maintain pressure.
- Too large → nasal mucosal
injury.
- Attach the nasal catheter to
the generator and temporarily cover with a finger to check that the
manometer shows the correct pressure (5 cm H₂O).
Connecting the Infant to the Device
- Wash hands or use alcohol rub.
- Ensure no obstruction in the
infant’s posterior nasal passages (congenital defect in some cases).
- Connect the infant to the
monitoring system (oxygen saturation and vital signs).
- Wear gloves.
- Suction the nose, mouth, and
pharynx using the largest catheter that can pass without resistance.
- Positioning: Raise the bed 30°, place a folded sheet under the
infant’s shoulders, slightly extend the neck to keep the airway open.
- Place the appropriately sized
cap, bending its edge 2–3 cm above the ears.
- Insert the double-prong nasal
catheter and secure with the cap’s attached strings, ensuring:
- Catheter fills both nostrils
completely.
- Skin around the nose is not
under tension.
- Catheter does not press on the
nasal septum.
- Space between the septum tip
and the catheter bridge.
- Catheter does not rest on the
upper lip.
- Secure the exhalation
corrugated tube to the forehead via cap straps.
- Insert a gastric feeding tube
for stomach suction, connecting it to a syringe to release excess air.
- Clean the upper lip and cheeks
with warm water, dry, and apply prescribed moisturizer to prevent pressure
sores.
Monitoring the Infant on the Device
- Check every 2–4 hours:
- Respiratory status: rate,
nasal flaring, chest retractions, breath sounds.
- Body temperature.
- Circulation: capillary refill,
blood pressure, heart rate.
- Neurological status: activity,
movement, reflexes.
- Gastrointestinal: abdominal
distention.
- Monitor devices for proper
function.
- Suction as needed (nose, mouth,
pharynx, stomach). Shorter suction intervals (<2 hours) may be needed
if:
- Increased respiratory effort.
- Higher oxygen requirements.
- Apnea or bradycardia episodes.
- Record volume and color of
secretions.
- Verify device function via
checklist.
- Reposition infant every 4–6
hours to prevent secretion accumulation.
- Drain and dispose of
condensation in tubing regularly.
Weaning from Nasal CPAP
Indicators for weaning:
- Overall improvement in the
infant’s condition.
- Spontaneous, comfortable
breathing.
- Reduced respiratory rate,
absence of chest retractions, improved color.
Weaning Steps:
- Gradually reduce oxygen
concentration by 2–5% at a time until reaching 21%, guided by pulse
oximetry or arterial blood gas analysis.
- Temporarily remove the nasal
catheter and observe:
- Increased breathing rate.
- Chest retractions.
- Decreased oxygen saturation.
- Apnea episodes.
- Cyanosis.
- If any signs appear → weaning
fails; resume CPAP and attempt again after 24 hours.
- Do not attempt weaning if
alveolar collapse is likely, as prevention is better than treatment.
Feeding During Nasal CPAP
- If stable, feed via the
gastrointestinal tract using gravity or syringe pump.
- Suction excess stomach air
before feeding.
Potential Complications of Nasal
CPAP
- Nasal septum injuries from
prolonged catheter pressure.
- Skin injuries from device
straps.
- Abdominal distention.
- Increased secretions.
- Pneumothorax.
Checklist for Neonatal NCPAP
Monitoring (Filled by Responsible Nurse Each Shift)

Checklist for Preparing the Neonatal
Continuous Positive Airway Pressure (NCPAP) System
|
Date ....../
....../ ......
|
Shift
------
|
Shift
------
|
Shift
------
|
Comments
|
|
Check
|
Yes / No
|
Yes / No
|
Yes / No
|
|
Nasal CPAP
device ready for use
|
|
|
|
|
|
Sufficient
number of connections and various sizes of nasal prongs available
|
|
|
|
|
|
Sterile
water bottles available
|
|
|
|
|
|
Oxygen
humidifier ready for use
|
|
|
|
|
|
Nurse’s
signature:
|
|
|
|
|